Title of article :
Simultaneous measurement of 89Sr and 90Sr in aqueous samples by liquid scintillation counting using the spectrum unfolding method
Author/Authors :
K. H. Hong، نويسنده , , Y. H. Cho، نويسنده , , Q. Meng and M. H. Lee ، نويسنده , , G. S. Choi، نويسنده , , C. W. Lee، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Abstract :
Two radiostrontium isotopes, 89Sr and 90Sr, were measured simultaneously by liquid scintillation counter coupling with a full spectrum DPM counting protocol. Strontium was separated from the aqueous media by extraction chromatography using Sr-spec resin. Individual count rates were evaluated from quench correlation curves for 89Sr and 90Sr related with SIS and tSIE quench indicators. The method exhibited a counting efficiency of 95% for 89Sr and 92% for 90Sr within their full energy channels and gave highly precise results when the sample was counted within 4 h after strontium separation. Lower limits of detection were 37 mBq/l for 90Sr and 32 mBq/l for 89Sr with a 60 min counting time.
Keywords :
Liquid scintillation counting , Full spectrum DPM , 90Sr , 89Sr
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes