Dosimetry of blood irradiation using an alanine/ESR dosimeter
Author/Authors :
F. Chen ، نويسنده , , D. T. Covas، نويسنده , , O. Baffa، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
4
From page :
13
To page :
16
Abstract :
A batch of 80 DL-alanine dosimeters was supplied to Hemocentro of the Hospital and Clinics of Faculdade de Medicina de Ribeirão Preto (HC-FMRP) SP, Brazil for the purpose of quality control of the radiation dose delivered to blood bags. The irradiation was made using two (40×40) cm2 parallel opposed radiation fields each with 80 cm of source to surface distance in the Radiotherapy Section of HC-FMRP with the 60Co teletherapy unit. The calculated radiation absorbed dose at the center of the box was 20 Gy. The dosimeter readings were performed using a Varian E-4 ESR Spectrometer operating in X-band. For the 80 dosimeters and over the irradiation volume throughout a blood bag, the minimum and maximum doses were 14 and 23 Gy, respectively. The mean dose was (18±2) Gy (1σ), and the coefficient of variability was 11.1%. Alanine dosimeters demonstrated easy handling, good precision and adequate sensitivity for this application.
Keywords :
Blood products , ESR , Alanine , quality control , Graft vs. host disease , dosimetry