Title of article
Analysis of a discrepancy in electron–beam dose comparison between chemical dosimeters and a calorimeter
Author/Authors
Zong Yuda، نويسنده , , Gao Juncheng، نويسنده , , TATSUO TABATA، نويسنده , , Zhang Yanli، نويسنده , , Yang Yuandi، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
4
From page
125
To page
128
Abstract
A 2.1% discrepancy was reported from an electron-beam dose comparison between liquid chemical dosimeters and a water calorimeter. Two methods, depth–dose measurement in a sandwich phantom and the use of a semiempirical depth–dose code EDMULT, were used to analyze the depth-dose distribution in the dosimeters and the calorimeter and to calculate their dose ratio. Results have shown that the discrepancy can be explained by the difference in the effective detection volume.
Keywords
Depth-dose distribution , absorbed dose , electron beam , Sandwich phantom , Semiempirical code