Title of article :
Characterization of three digital signal processor systems used in gamma ray spectrometry
Author/Authors :
N. Reguigui، نويسنده , , J. Morel Symons، نويسنده , , H. Ben Kraiem، نويسنده , , Zakwan A. Mahjoub، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
5
From page :
93
To page :
97
Abstract :
Various manufacturers have recently introduced digital signal processing systems that allow data acquisition in gamma spectrometry at high-input counting rates (several thousand pulses per second). In these systems, the signal digitization is performed immediately following the preamplification stage. This allows digital shaping and filtering of the signal which increases the number of possible combinations in signal shaping and as a consequence, optimizes the resolution as a function of the detector characteristics and the counting rate. Basic characteristic parameters of three digital signal processors that were recently introduced in the market have been studied and compared to those of an analog system. This study is carried out using a hyper-pure coaxial type germanium detector and 57Co, 60Co and 137Cs radioactive sources. Performance parameters such as energy resolution, system throughput, and counting losses that are due to dead time and pile-up effects are presented and discussed.
Journal title :
Applied Radiation and Isotopes
Serial Year :
2002
Journal title :
Applied Radiation and Isotopes
Record number :
541131
Link To Document :
بازگشت