Title of article :
A dual-compensated cryogenic microcalorimeter for radioactivity standardizations
Author/Authors :
R. Colle، نويسنده , , B. E. Zimmerman، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
Efforts are underway by our laboratory to develop a microcalorimeter that can be routinely used for radioactivity standardizations of nuclides that decay by pure β-emission or by low-Z electron capture. The prototype calorimeter consists of a cryostat with two temperature-controlled stages and a base stage that are operated at nominal temperatures of 8 K. A unique aspect of the calorimeterʹs design is the ability to repeatedly engage and disengage the radioactive source from the second stage heat path using a magnetically activated elevator. The measurement of the total power from a radioactive sample is obtained from the difference in the second stage power with and without the source in place. As a result of extensive performance evaluations using 90Sr–90Y and 32P brachytherapy seeds as well as with an internal calibration heater, many initial design flaws have been identified and are being addressed.
Keywords :
Measurement , Instrument , Microcalorimetry , radioactivity , Standards , Brachytherapy , Calorimetry , cryogenics , Metrology
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes