• Title of article

    A new absolute method for the standardization of radionuclides emitting low-energy radiation

  • Author/Authors

    E. Leblanc، نويسنده , , P. de Marcillac، نويسنده , , N. Coron، نويسنده , , J. Leblanc، نويسنده , , M. Loidl، نويسنده , , J. F. Metge، نويسنده , , J. Bouchard، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    7
  • From page
    245
  • To page
    251
  • Abstract
    Microcalorimeters (or bolometers) operated at temperatures below 100 mK allow individual counting of photons and electrons with a very low energy detection threshold. The physics is based on the pulse temperature increase of the target (or absorber) of the detector due to the complete absorption of both electrons and photons. Since this target can be constructed with a perfect 4-π geometry, a bolometer offers potentially a new method for absolute activity measurements of radionuclides emitting low-energy radiation. In this paper we present our first results of a feasibility study of activity standardization of a 55Fe solution with a prototype 4-π bolometer.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2002
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    541154