Title of article :
Comparison of lead attenuation and lead hardening equivalence of materials used in respect of diagnostic X-ray shielding
Author/Authors :
Akintunde Akangbe Okunade، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
6
From page :
819
To page :
824
Abstract :
Present interest is in the shielding of diagnostic X-ray units. Numerical comparison has been made of the attenuation and hardening properties of lead and some particular alternative materials: steel, plate glass and gypsum wallboard. Results show, for particular choices of thickness, that lead and steel can be made to provide closely similar attenuation and spectral hardening, values of lead attenuation equivalent (LAE) and lead hardening equivalent (LHE) thicknesses being nearly the same. Significant differences in the attenuation and hardening properties of lead are found in comparison with plate glass and gypsum wallboard. LAE produces better matching of exposure for lead-plate glass and lead-gypsum wallboard than LHE.
Keywords :
Diagnostic X-rays , Shielding , attenuation , Lead hardening equivalence
Journal title :
Applied Radiation and Isotopes
Serial Year :
2002
Journal title :
Applied Radiation and Isotopes
Record number :
541371
Link To Document :
بازگشت