• Title of article

    Response characteristics of a tandem ionization chamber in standard X-ray beams

  • Author/Authors

    Alessandro M. Costa، نويسنده , , Linda V. E. Caldas، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    6
  • From page
    495
  • To page
    500
  • Abstract
    A tandem ionization chamber was developed for quality control programs of the kilovoltage X-ray equipment used in radiotherapy. This chamber was tested and calibrated in low-energy X-ray beams in accordance with international recommendations, and it showed a satisfactory level of performance, mainly with regard to its main use, namely formation of a tandem system for confirmation of half-value layers in X-radiation beams previously determined by the conventional method. In this developed system no absorbers or special setups are necessary.
  • Keywords
    radiotherapy , Tandem system , quality control , X-radiation , Ionization chambers