Abstract
A tandem ionization chamber was developed for quality control programs of the kilovoltage X-ray equipment used in radiotherapy. This chamber was tested and calibrated in low-energy X-ray beams in accordance with international recommendations, and it showed a satisfactory level of performance, mainly with regard to its main use, namely formation of a tandem system for confirmation of half-value layers in X-radiation beams previously determined by the conventional method. In this developed system no absorbers or special setups are necessary.
Keywords
radiotherapy , Tandem system , quality control , X-radiation , Ionization chambers