Title of article
A 3D high-resolution gamma camera for radiopharmaceutical studies with small animals
Author/Authors
G. K. Loudos، نويسنده , , S. Koulouridis and K. S. Nikita، نويسنده , , N. D. Giokaris، نويسنده , , E. Styliaris، نويسنده , , S. C. Archimandritis، نويسنده , , A. D. Varvarigou، نويسنده , , C. N. Papanicolas، نويسنده , , S. Majewski، نويسنده , , D. Weisenberger، نويسنده , , R. Pani، نويسنده , , F. Scopinaro، نويسنده , , N. K. Uzunoglu، نويسنده , , D. Maintas، نويسنده , , K. Stefanis، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
8
From page
501
To page
508
Abstract
The results of studies conducted with a small field of view tomographic gamma camera based on a Position Sensitive Photomultiplier Tube are reported. The system has been used for the evaluation of radiopharmaceuticals in small animals. Phantom studies have shown a spatial resolution of 2 mm in planar and 2–3 mm in tomographic imaging. Imaging studies in mice have been carried out both in 2D and 3D. Conventional radiopharmaceuticals have been used and the results have been compared with images from a clinically used system.
Keywords
SPECT , High resolution , Radiopharmaceutical studies , Small animal imaging , PSPMT