• Title of article

    A review of the triple-to-double coincidence ratio (TDCR) method for standardizing radionuclides

  • Author/Authors

    R. Broda، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    10
  • From page
    585
  • To page
    594
  • Abstract
    This paper gives a review of the development of the triple-to-double coincidence ratio (TDCR) method for standardizing pure β- and pure electron capture (EC)-radionuclides. The initial idea of Schwerdtel and practical realization and elaboration of the TDCR method by Pochwalski is presented. The main aspects and theoretical principle of the method are summarized and progress on the development of the theoretical model is presented. The details of the activity determination of pure β- and pure EC radionuclides are discussed, along with specific details in the application of the method in each case. The enhanced TDCR method based on the MAC3 electronic system developed in France is also discussed. Different applications of the TDCR method in metrology laboratories around the world are also pointed out.
  • Keywords
    Liquid scintillator counter , TDCR method , Coincidence , radioactivity , counting efficiency , radionuclide
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2003
  • Journal title
    Applied Radiation and Isotopes
  • Record number