    Performance of CdTe, HPGe and NaI(Tl) detectors for radioactivity measurements

  • Author/Authors

    A. Perez-Andujar، نويسنده , , L. Pibida، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    7
  • From page
    41
  • To page
    47
  • Abstract
    The characteristics and performances of the NIST High-Purity Germanium (HPGe), Sodium Iodide (NaI(Tl)) and Cadmium Telluride (CdTe) gamma-ray detectors were studied. The efficiencies, the minimum detectable activities and the energy resolutions of each were measured and compared. The data were analyzed using different software packages. The measurements were performed in a low background radiation environment using calibrated point sources. The CdTe detector showed a higher energy resolution than the NaI(Tl) detector but a lower efficiency than both the HPGe and NaI(Tl) detectors. This makes it a suitable detector only for relatively high activity measurements restricted to the 35 keV to 200 keV energy range.
  • Keywords
    Gamma-ray spectrometry , CdTe detector , NaI(Tl) detector , HPGe detector