Title of article
Establishing equivalence for activity standards of short-lived radionuclides using the NPL secondary standard radionuclide calibrator
Author/Authors
M. J. Woods، نويسنده , , M. Baker، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
6
From page
499
To page
504
Abstract
Conventional comparison techniques used between National Metrology Institutes are not practicable for short-lived radionuclides because of geographical separations and transport difficulties. The NPL Secondary Standard Radionuclide Calibrator provides an alternative approach and a comparison was conducted with 18F to investigate its feasibility. The exercise was successful and the paper details the protocol used, the quality assurance mechanisms introduced to underpin the comparison and an analysis of the results. It was also demonstrated that this approach could be linked to the BIPM SIR system. Recommendations are presented for the extension of this work to other suitable, short-lived radionuclides.
Keywords
equivalence , NPL secondary standard radionuclide calibrator , 18F , radioactivity , Comparisons
Journal title
Applied Radiation and Isotopes
Serial Year
2004
Journal title
Applied Radiation and Isotopes
Record number
541634
Link To Document