• Title of article

    A semi-empirical approach for determination of low-energy gamma-emmiters in sediment samples with coaxial Ge-detectors

  • Author/Authors

    E. G. San Miguel، نويسنده , , J. P. Perez-Moreno، نويسنده , , J. P. Bolivar، نويسنده , , R. Garcia-Tenorio، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    361
  • To page
    366
  • Abstract
    A simple method for the determination in sediment samples of low-energy gamma-emitters with coaxial Ge detectors and using a cylindrical sample geometry is outlined. This method allows an easy estimation of self-absorption effects by knowing the apparent densities and the composition of the investigated samples. As calibration matrixes solid samples enriched in known amounts of natural radionuclides emitting low-energy gamma rays were used. In this way, the difficult homogenisation steps needed when liquid spikes are added to a solid blank for manufacturing the calibration sample are avoided. The method has been carefully checked and validated by applying it to sediment samples with known activities of some low-energy gamma-ray emitters. Additionally, these sediment measurements have allowed to evaluate the magnitude and importance of the self-absorption corrections
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2004
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    541761