• Title of article

    The TL analysis methods used to determine absorbed gamma doses in vivo for the BNCT patients treated at FiR 1

  • Author/Authors

    J. Karila، نويسنده , , T. Sepp?l?، نويسنده , , T. Serén، نويسنده , , P. Kotiluoto، نويسنده , , I. Auterinen، نويسنده , , C. Aschan، نويسنده , , L. Kankaanranta، نويسنده , , S. Savolainen، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    5
  • From page
    793
  • To page
    797
  • Abstract
    The gamma dose determination using thermoluminescent (TL) dosimeters in mixed neutron-gamma fields, such as in boron neutron capture therapy (BNCT), is difficult due to the thermal neutron sensitivity of the detectors; especially when equipment capable of glow curve analysis is not available. The two TL analysis methods used previously in Finnish BNCT to correct the measured TL signal to obtain absorbed gamma dose in vivo were studied and compared, and an enhanced method was introduced. The three TL methods were found surprisingly consistent despite, e.g. the rough estimate made in the first method.
  • Keywords
    TL dosimeter , Gamma dose , BNCT , TPS , In Vivo
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2004
  • Journal title
    Applied Radiation and Isotopes
  • Record number