Title of article :
Enhanced blood boron concentration estimation for BPA-F mediated BNCT
Author/Authors :
M. Kortesniemi، نويسنده , , T. Sepp?l?، نويسنده , , I. Auterinen، نويسنده , , S. Savolainen، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
5
From page :
823
To page :
827
Abstract :
The blood boron concentration regulates directly the BNCT irradiation time in which the prescribed dose to the patient is delivered. Therefore a proper estimation of the blood boron concentration for the treatment field based on the measured blood samples before irradiation is required. The bi-exponential model fit using Levenberg—Marquardt method was implemented for this purpose to provide the blood boron concentration estimates directly to the treatment data flow during the BNCT procedure. The harmonic mean bi-exponential decay half-lives of the studied patient data (n=28) were 15±8 and 320±70 min for the faster and slower half-life. The model uncertainty (n=28) was reasonably low, 0.7±0.1 μg/g (about 5%). The implemented algorithm provides a robust method for temporal blood boron concentration estimation for BPA-F mediated BNCT. Utilization of the infusion data improves the reliability of the estimate. The overall data flow during the treatment fulfills the practical requirements concerning the BNCT procedure.
Keywords :
Boronophenylalanine , fructose , BPA–F , Pharmacokinetics , Levenberg–Marquardt
Journal title :
Applied Radiation and Isotopes
Serial Year :
2004
Journal title :
Applied Radiation and Isotopes
Record number :
541832
Link To Document :
بازگشت