Title of article
Dose evaluation and measurement of the 188Re liquid-filled balloon in intravascular brachytherapy
Author/Authors
Uei-Tyng Lin، نويسنده , , Chien-Hau Chu، نويسنده , , Bor-Tsung Hsieh، نويسنده , , Wen-Song Hwang، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
11
From page
1323
To page
1333
Abstract
We compared the Monte Carlo evaluation and GafChromic MD-55 film experimental measurement of 188Re liquid-filled balloons in intravascular brachytherapy using two phantoms of 6 mm vascular diameter, a phantom of 4 mm vascular diameter and a phantom of 3 mm vascular diameter. A dose-evaluation interface program was developed and was shown to be capable of quickly providing information such as the necessary 188Re source irradiation time to deliver a prescribed dose.
Keywords
restenosis , brachytherapy , Monte Carlo , Rhenium-188
Journal title
Applied Radiation and Isotopes
Serial Year
2004
Journal title
Applied Radiation and Isotopes
Record number
541913