Title of article :
Dose evaluation and measurement of the 188Re liquid-filled balloon in intravascular brachytherapy
Author/Authors :
Uei-Tyng Lin، نويسنده , , Chien-Hau Chu، نويسنده , , Bor-Tsung Hsieh، نويسنده , , Wen-Song Hwang، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
11
From page :
1323
To page :
1333
Abstract :
We compared the Monte Carlo evaluation and GafChromic MD-55 film experimental measurement of 188Re liquid-filled balloons in intravascular brachytherapy using two phantoms of 6 mm vascular diameter, a phantom of 4 mm vascular diameter and a phantom of 3 mm vascular diameter. A dose-evaluation interface program was developed and was shown to be capable of quickly providing information such as the necessary 188Re source irradiation time to deliver a prescribed dose.
Keywords :
restenosis , brachytherapy , Monte Carlo , Rhenium-188
Journal title :
Applied Radiation and Isotopes
Serial Year :
2004
Journal title :
Applied Radiation and Isotopes
Record number :
541913
Link To Document :
بازگشت