Title of article
Proposed definitions for isodose flatness and symmetry in clinical radiotherapy beams
Author/Authors
Eduardo Galiano، نويسنده , , Tara Joly، نويسنده , , Federico Wiebe، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
6
From page
1361
To page
1366
Abstract
In clinical radiotherapy it is important that beam intensity be as homogeneous as possible to reduce the probability of treatment failure. As an extension of the well established concepts of beam flatness and symmetry for characterizing radiotherapy beams, the concepts of isodose flatness and symmetry are introduced. The definitions are tested with actual data obtained from a Co-60 unit and a linear accelerator and results are presented. The concepts as defined appear to provide a useful quantitative indication of beam homogeneity.
Keywords
Radiotherapy beams , Beam flatness , Beam symmetry , Isodose flatness , Isodose symmetry
Journal title
Applied Radiation and Isotopes
Serial Year
2004
Journal title
Applied Radiation and Isotopes
Record number
541919
Link To Document