Title of article
The radiation field characteristics of a 137Cs source used for calibration of radiation protection instruments
Author/Authors
E. Fernandes، نويسنده , , D. Freire، نويسنده , , A. C. de Freitas، نويسنده , , C. E. deAlmeida، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
6
From page
1425
To page
1430
Abstract
In this work, an experimental study of the radiation beam characteristics of the photon beams produced by commercial irradiator that uses a single collimated 137Cs source is performed. A set of lead attenuators are placed at the exit window of the irradiator to vary the air Kerma that is required to cover the instrument scales at a particular calibration distance. The results strongly enhance the findings of previous studies that a single source with appropriate attenuation and collimation is a practical solution for calibration laboratories with an acceptable overall accuracy.
Keywords
ionization chambers , Calibration , Radiation detector , Air-kerma rate , 137Cs , Lead absorbers and radiation field characterization , dosimetry