• Title of article

    Use of alanine for dosimetry intercomparisons among Italian radiotherapy centers

  • Author/Authors

    C. De Angelis، نويسنده , , V. De Coste، نويسنده , , P. Fattibene، نويسنده , , S. Onori، نويسنده , , E. Petetti، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    261
  • To page
    265
  • Abstract
    A pilot program of postal dosimetry intercomparison among 16 Italian Radiotherapy Centers was performed using the ISS Alanine/EPR dosimetry as a transfer system. Dosimeters were irradiated at 10 Gy with high-energy photon beams, both in reference condition in a water phantom and in an anthropomorphic phantom during the simulated treatment of rectum cancer. Intercomparison design along with alanine performances analyzing the different contributions to the combined uncertainty in dose assessment are reported. Main results of the pilot intercomparison, terminated in 2002, are also presented.
  • Keywords
    Alanine dosimetry , Quality assurance in radiotherapy , Dosimetry intercomparison
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2005
  • Journal title
    Applied Radiation and Isotopes
  • Record number