Use of the entire spectrum of irradiated alanine for dosimetry
Author/Authors :
J.M. Dolo، نويسنده , , F. Moignau، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
Alanine is an amino acid commonly used in ESR dosimetry as a reference detector. The classic approach for the measurement of irradiated samples is to determine the amplitude of the central peak of the first derivative spectrum. It is generally considered that this technique represents the best and most reproducible solution for achieving an accurate proportionality between the concentration of free radicals inside the resonant cavity, characterized by the amplitude, and the dose. It is also accepted that this central peak corresponds to the free radical CH3.CHCOO−. The hyperfine structure of this radical in the spectrum shows five main peaks with the approximate ratios 1:4:6:4:1 as regards coupling. This paper presents another approach featuring analysis of the entire spectrum: (i) ratios of identified peaks, (ii) ratio variation vs time with regard to several parameters affecting fading. These variations in the alanine spectrum are probably correlated with the variation of the concentrations of different free radical species. These variations and their positions in the spectrum are very important constraints that increase the uncertainty of this type of measurement.
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes