Title of article :
Release of foreign bodies (particles) by clinical use of intravenous infusion sets
Author/Authors :
Henrik Madsen، نويسنده , , Ole Winding، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
4
From page :
663
To page :
666
Abstract :
In clinical practice, stripping the plastic tubes of intravenous (i.v.) infusion sets with a scissor blade is a commonly used method for re-establishing flow in malfunctioning i.v. sets. The present investigation concludes that this procedure results in release of plastic particles from the luminal wall of the tube. Particles are subsequently flushed into the patient. The average amount of particles released exclusively from the i.v. infusion sets under these circumstances may exceed the standards for acceptable particle content per millilitre in large volume i.v. injectable fluids, according to the British Pharmacopoeia (1980) and the United States Pharmacopoeia (1990).
Keywords :
Intravenous infusion sets , contamination , Particles , Stripping
Journal title :
Biomaterials
Serial Year :
1996
Journal title :
Biomaterials
Record number :
542442
Link To Document :
بازگشت