Title of article
Characterization of corrosion products on a copper-containing intrauterine device during storage at room temperature
Author/Authors
J. M. Bastidas، نويسنده , , J. Simancas، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
4
From page
247
To page
250
Abstract
This paper studies the characterization of corrosion products formed on corroded and uncorroded copper-containing intrauterine devices stored at room temperature for a period of 30 months. The experimental techniques used were X-ray photo-electron spectroscopy and Auger electron spectroscopy. The compounds found were cuprite (Cu2O) and tenorite (CuO). The latter was the main compound formed on corroded samples, forming thin tarnish films.
Keywords
corrosion products , copper , Intrauterine contraceptive device , X-ray photo-electron spectroscopy , Auger electron spectroscopy
Journal title
Biomaterials
Serial Year
1997
Journal title
Biomaterials
Record number
542707