Title of article :
Surface-dimpled commercially pure titanium implant and bone ingrowth
Author/Authors :
Jianguo Li، نويسنده , , Hailhong Liao، نويسنده , , Bahaman Fartash، نويسنده , , Leif Hermansson، نويسنده , , Thomas Johnsson، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
6
From page :
691
To page :
696
Abstract :
The purpose of this investigation was to examine the effect of the surface macrostructure of a dimpled commercially pure titanium (cp Ti) implant on bone ingrowth in vivo by means of histological examination and a push-out test. Cylindrical implants were inserted in one femur of each experimental rabbit and the animals were killed at 1.5, 3 and 13 months after implantation. The femur with the implant of each animal was then examined in a push-out test. The fracture surfaces of the bone-implant interface after the push-out test were examined under light and electron microscopy. It seems that the dimpled cp Ti surface results in the increased retention of the cp Ti implant in bone due to interlocking between vital bone and the dimples.
Keywords :
bone formation , fracture interface , Shear strength , surface macrostructure
Journal title :
Biomaterials
Serial Year :
1997
Journal title :
Biomaterials
Record number :
542768
Link To Document :
بازگشت