• Title of article

    Measurement and evaluation of galvanic corrosion between titanium/Ti6Al4V implants and dental alloys by electrochemical techniques and auger spectrometry

  • Author/Authors

    Brigitte Grosgogeat، نويسنده , , Lucien Reclaru، نويسنده , , Michèle Lissac، نويسنده , , Francis Dalard، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    9
  • From page
    933
  • To page
    941
  • Abstract
    The purpose of this study was to investigate, in different experimental conditions, the galvanic corrosion phenomena which can exist between a dental suprastructure and a dental implant. The electrochemical behavior of 7 alloy superstructures with titanium and titanium alloy (Ti6Al4V) implants was investigated by electrochemical means in Fusayama–Meyer de-aerated saliva and Carter–Brugirard (AFNOR) non de-aerated saliva. Different techniques were used to obtain the value of the galvanic coupling current and potential for each couple. All showed very low corrosion rates, ranging from 10-6 to 10-8 A. Surface analysis confirmed these results.
  • Journal title
    Biomaterials
  • Serial Year
    1999
  • Journal title
    Biomaterials
  • Record number