Abstract
The corrosion behavior of surgical implant CoCr alloy and REX 734 steel has been investigated. The pitting or crevice corrosion potentials have been determined to reach values as high as 500 mV vs. SCE for CoCr and 450 mV vs. SCE for REX 734. The galvanic corrosion behavior of CoCr/REX 734 couples has been evaluated with various electrochemical techniques. The measurement of the corrosion current of the galvanic couple as well as its prediction by applying mixed potential theories on measured potentiodynamic polarization curves revealed low galvanic currents in the range of nanoamperes.
Keywords
Electrochemistry , pitting , Crevice and galvanic corrosion , stainless steels , CoCr alloy , Orthopedic prostheses