Title of article :
Collapse pressures of biodegradable stents
Author/Authors :
Subbu Venkatraman، نويسنده , , Tan Lay Poh، نويسنده , , Tjong Vinalia، نويسنده , , Koon Hou Mak، نويسنده , , Freddy Boey، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
7
From page :
2105
To page :
2111
Abstract :
Biodegradable stent prototypes were produced from poly -lactic acid polymers with different molecular weights. The effects of molecular weight, drug incorporation and stent design on the collapse pressure of the stents were evaluated. While molecular weights did not show a significant effect on the collapse pressure of the stents, drug incorporation at high percentage decreased the collapse pressure of the stents substantially. Cryogenic fracture surfaces showed significant drug agglomeration as the concentration increased. The design of the stent was also found to a have significant effect on the collapse pressure. The stent produced from the same material has a higher collapse pressure when the load bearing surface area is increased.
Keywords :
coronary stents , biodegradable , mechanical properties , helical , PLLA