The influence of fluorohydroxyapatite (FHA) coating and surface roughness of Ti6Al4V implants on bone response was investigated. Uncoated and FHA-coated screws with lower (LR and LR+FHA; Ra: 5.7±0.2 μm) and higher (HR and HR+FHA; Ra: 21.8±0.9 μm) surface roughness, were inserted into the diaphyses of 8 sheep tibiae. Twelve weeks after implantation, extraction torque and bone-to-implant contact were evaluated. The smoothest surfaces showed an improved extraction torque and significant differences were observed between LR and HR (−24.6%, p<0.0005), LR and HR+FHA (−30.7%, p<0.0005), LR+FHA and HR (−17.4%, p<0.005), and LR+FHA and HR+FHA (−24.0%, p<0.005). The bone-to-implant contact data paralleled the biomechanical data: the smoother the surface, the greater the bone-to-implant contact. Significant (p<0.0005) decreases in bone-to-implant contact were observed between LR+FHA and HR (−24.2%), and between LR+FHA and HR+FHA (−29.2%). The current findings suggest that LR surfaces significantly improve the osteointegration rate of implanted cortical screws independently of the FHA coating.
Keywords :
Coating , Titanium , Fluorohydroxyapatite , Roughness , Osteointegration
Journal title :
Biomaterials
Journal title :
Biomaterials