• Title of article

    Surface modifications induced by ns and sub-ps excimer laser pulses on titanium implant material

  • Author/Authors

    M. Bereznai، نويسنده , , I. Pels?czi، نويسنده , , Z. T?th، نويسنده , , K. Turz?، نويسنده , , M. Radnai، نويسنده , , Z. Bor، نويسنده , , A. Fazekas، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    7
  • From page
    4197
  • To page
    4203
  • Abstract
    Medical implants used in oral and orthopaedic surgery are mainly produced from titanium. Their biological behaviour, e.g. osseointegration, essentially depends on both the chemical composition and the morphology of the surface. Modifications achieved by excimer laser irradiation of titanium samples were investigated in order to improve their surface characteristics so as to facilitate biointegration. To enlarge the effective interfacial area of bone–implant contact, holes were ablated by laser pulses of ns or sub-ps length. During ns ablation, crown-like projecting rims formed around the borders of the holes. Ultra-short (0.5 ps) KrF excimer laser pulses were successfully applied to avoid these undesirable formations. Since a smooth dental implant surface is necessary to maintain a healthy connection with the soft tissues, laser polishing of samples was investigated, too. Irradiation with a series of ns laser pulses resulted in effective smoothing, as measured with atomic force microscope. X-ray photoelectron spectroscopy analysis of the laser-polished titanium surface revealed that laser treatment led to a decrease of the surface contamination and in thickening of the oxide layer. X-ray diffraction measurements demonstrated that the original α-titanium crystal structure was preserved.
  • Keywords
    Osseointegration , Surface modification , Laser ablation , surface roughness , Titanium
  • Journal title
    Biomaterials
  • Serial Year
    2003
  • Journal title
    Biomaterials
  • Record number

    545063