Title of article
Increased viable osteoblast density in the presence of nanophase compared to conventional alumina and titania particles
Author/Authors
Luke G. Gutwein ، نويسنده , , Thomas J. Webster، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
9
From page
4175
To page
4183
Abstract
In the present in vitro study, osteoblast (bone-forming cells) viability and cell density were investigated when cultured in the presence of nanophase compared to conventional (i.e. micron) alumina and titania particles at various concentrations (from 10,000 to 100 μg/ml of cell culture media) for up to 6 h. Results confirmed previous studies of the detrimental influences of all ceramic particulates on osteoblast viability and cell densities. For the first time, however, results provided evidence of increased apoptotic cell death when cultured in the presence of conventional compared to nanophase alumina and titania particles. Moreover, since material characterization studies revealed that the only difference between respective ceramic particles was nanometer- and conventional-dimensions (specifically, phase and chemical properties were similar between respective nanophase and conventional alumina as well as titania particles), these results indicated that osteoblast viability and densities were influenced solely by particle size. Such nanometer particulate wear debris may result from friction between articulating components of orthopedic implants composed of novel nanophase ceramic materials. Results of a less detrimental effect of nanometer—as compared to conventional-dimensioned particles on the functions of osteoblasts provide additional evidence that nanophase ceramics may become the next generation of bone prosthetic materials with increased efficacy and, thus, deserve further testing.
Keywords
orthopedic , wear debris , Titania , Nanoparticle , Osteoblast , Apoptosis , alumina
Journal title
Biomaterials
Serial Year
2004
Journal title
Biomaterials
Record number
545608
Link To Document