• Title of article

    Loss of ions in cavity ionization chambers

  • Author/Authors

    N. Takata، نويسنده , , N.T. Tran، نويسنده , , E. Kim، نويسنده , , P. Marsoem، نويسنده , , T. Kurosawa، نويسنده , , Y. Koyama، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    4
  • From page
    805
  • To page
    808
  • Abstract
    Ion losses due to initial recombination, volume recombination, and back diffusion were each determined by measurements and calculations for different size cylindrical ionization chambers and spherical ionization chambers. By measuring signal currents from these ionization chambers irradiated with 60Co gamma rays, two groups of ion losses were obtained. (Group 1) Ion loss due to initial recombination and diffusion, which changes proportionally to the inverse of the voltage applied to the ionization chambers; (and group 2) ion loss due to volume recombination, which changes proportionally to the inverse of the square of the applied voltage. The diffusion loss was obtained separately by computing electric field distributions in the ionization chambers. It was found that diffusion loss is larger than initial recombination loss for the cylindrical ionization chambers and vise versa for the spherical ionization chambers.
  • Keywords
    ionization chamber , recombination , diffusion , Signal current , Volume , Initial , Ion loss
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2005
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    547854