    Close-geometry efficiency calibration of p-type HPGe detectors with a Cs-134 point source

  • Author/Authors

    P. DeFelice، نويسنده , , A. Fazio، نويسنده , , T. Vidmar، نويسنده , , M. Korun، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    4
  • From page
    1303
  • To page
    1306
  • Abstract
    When close-geometry detector calibration is required in gamma-ray spectrometry, single-line emitters are usually used in order to avoid true coincidence summing effects. We managed to overcome this limitation by developing a method for the determination of the efficiency of p-type HPGe detectors in close-geometry with a calibrated Cs-134 point source. No separate determination of coincidence summing correction factors is required and a single measurement furnishes the full-energy-peak efficiencies in the 475–1365 keV energy range.
  • Keywords
    True coincidence summing , Close geometry , Detector calibration , Gamma-ray spectrometry , Cs-134