Application of PENELOPE code to the efficiency calibration of coaxial germanium detectors
Author/Authors :
M. Jurado Vargas، نويسنده , , A.L. Guerra، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
The code PENELOPE is applied here to the efficiency calibration exercise described in the EUROMET project 428 [Lépy et al., 2001. Appl. Radiat. Isot. 55, 493]. We evaluated the peak efficiencies for a coaxial HPGe detector in the range 60–2000 keV, for point sources located at various distances from the detector and for a cylindrical box containing two different matrixes: a low-density silica and a hydrochloric acid solution. After an optimization of several detector parameters, the results were in good agreement with the experimental values.
Keywords :
Efficiency calibration , Monte Carlo simulation , Gamma-ray spectrometry , Ge detectors
Journal title :
Applied Radiation and Isotopes