Abstract :
An industry grade dual energy X-ray absorptiometry (DEXA) scanner was calibrated for Wool Base determination. The calibration used 201 Crossbred and Merino wool samples, and a further 72 samples to validate the calibration. The prediction correlation had the smallest residual standard deviation (RSD) when the independently measured mean fibre diameter (MFD) was included in the multiple regression analysis. Best results were achieved when separate calibrations were used for individual wool breeds. The RSD for the Merino calibration set of 44 samples was 1.88, when the MFD was included in the regression, and 2.1 without. The RSD for 144 Crossbred samples was 1.73 including the MFD, and 2.59 without. The validation trial with 46 Crossbred and 24 Merino wool samples resulted in RSD of 2.35 and 2.23, respectively. An excellent DEXA repeatability was achieved at a standard deviation of not, vert, similar0.2%. Improvement of the calibration is expected from concurrent laboratory testing and scanning. The research shows the promising potential for DEXA as a tool to determine Wool Base.
Keywords :
DEXA , Wool base , X-ray scanning , DXA , Measurement techniques , Non-invasive evaluation