Title of article :
Calibration of a 137Cs γ-ray beam irradiator using large size chambers
Author/Authors :
R. Minniti، نويسنده , , S.M. Seltzer، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
A 137Cs γ-ray beam irradiator has been calibrated in terms of air kerma using large size chambers. The available air-kerma rates range between 1.8 μGy/h (0.2 mR/h) and 5.3 mGy/h (0.6 R/h). Large-volume chambers were used to characterize the source in terms of the radiation quantity air kerma (and exposure). Two types of chambers with significantly different characteristics and energy responses were used. This work shows that very good agreement can be obtained between the measurements performed with such different types of chambers. An agreement of 0.3% is observed between chambers even for the lowest air-kerma rates measured.
Keywords :
Calibration coefficient , ionization chamber , Cs-137 , ?-Ray , Secondary standard , Scattered radiation , Air kerma
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes