    We have developed a fully automatic method for the synthesis of 16α-[18F]fluoroestradiol ([18F]FES) using a disposable cassette system and conventional [18F]FDG module. [18F]FES was synthesized using a GE TracerLab MX module and a modified module control

  • Author/Authors

    C.W. Chang، نويسنده , , T.K. Chou، نويسنده , , R.S. Liu، نويسنده , , S.J. Wang، نويسنده , , W.J. Lin، نويسنده , , C.H. Chen، نويسنده , , H.E. Wang، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    5
  • From page
    682
  • To page
    686
  • Abstract
    This study aimed to develop an automated synthesis of [18F]fluoromisonidazole ([18F]FMISO) using a Scanditronix Anatech RB III robotic system. [18F]HF was produced via the 18O(p,n)18F reaction using a Scanditronix MC17F cyclotron. On average, a typical run produced [18F]FMISO with an uncorrected radiochemical yield of 30±5% at end of synthesis (EOS) from the irradiation of 95% enriched [18O]water. The total synthesis time was 65 min. The retention time of [18F]FMISO (the radio-peak) was 4.9 min, which was consistent with the authentic FMISO (the ultraviolet peak). The radiochemical purity was greater than 97%. Preparation of [18F]FMISO using the automated robotic system is highly reliable and reproducible, and the radiation burden for the operator can be largely reduced. Sufficient radioactivities of [18F]FMISO could be obtained for non-invasive tumor hypoxia imaging in vivo with positron emission tomography (PET).