Title of article
Determination of uranium in aqueous attenuating samples using gamma-ray spectrometry
Author/Authors
Chhavi Agarwal، نويسنده , , P.C. Kalsi، نويسنده , , A. Mhatre، نويسنده , , A. Goswami، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
3
From page
1386
To page
1388
Abstract
In the present work, a method for determination of uranium concentration in aqueous solution in standard geometry from attenuating samples has been developed based on modification of the empirical approach of Venkataraman and Croft [2003. Determination of plutonium mass using gamma-ray spectrometry. Nucl. Instrum. Methods Phys. Res. A 505, 527–530]. The method makes use of the multiple gamma (γ)-rays emitted by 235U and depends on the empirical relation between apparent mass of the sample and γ-ray energy. It was possible to determine uranium concentration in the range of 12–400 mg/ml rapidly by this method without applying transmission corrections.
Keywords
Uranium assay , AQUEOUS SOLUTIONS , Gamma spectrometry
Journal title
Applied Radiation and Isotopes
Serial Year
2007
Journal title
Applied Radiation and Isotopes
Record number
548331