Title of article :
A comparison of the performance characteristics of four film dosimeters in a 10-MeV electron beam
Author/Authors :
Zofia Peimel-Stuglik، نويسنده , , S?awomir Fabisiak، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
7
From page :
346
To page :
352
Abstract :
Performance characteristics of four routine film dosimeters (CTA, B-3, PVC and LiF (Sunna)) in a 10-MeV electron beam were investigated. Dose response curves for the range 5–40 kGy are presented. The dose response curve is linear for the CTA film, supralinear for the LiF (Sunna) film and sublinear for the B-3 and PVC films. The low sensitivity of the CTA film to ionizing radiation restricts its use to high doses, while the supralinear response curve of the LiF (Sunna) dosimeter allows this film to be used only in measurements of low doses. The optical absorption temperature coefficients for the CTA, B-3 and LiF (Sunna) films were found to be at the level of ±0.1%/ring operatorC. The temperature coefficient for the PVC film decreased from +0.35%/ring operatorC immediately after irradiation to +0.04%/ring operatorC three months later. Because of the significant time dependence of the PVC film signal, a careful standardization of the heating regimen and of the time between the irradiations and measurements is necessary.
Keywords :
Film dosimeters , PVC , B-3 , Sunna , electron beam , CTA , LiF
Journal title :
Applied Radiation and Isotopes
Serial Year :
2008
Journal title :
Applied Radiation and Isotopes
Record number :
548384