Title of article
CREPT-MCNP code for efficiency calibration of HPGe detectors with the representative point method
Author/Authors
Jun Saegusa، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
6
From page
774
To page
779
Abstract
The representative point method for the efficiency calibration of volume samples has been previously proposed. For smoothly implementing the method, a calculation code named CREPT-MCNP has been developed. The code estimates the position of a representative point which is intrinsic to each shape of volume sample. The self-absorption correction factors are also given to make correction on the efficiencies measured at the representative point with a standard point source. Features of the CREPT-MCNP code are presented.
Keywords
CREPT-MCNP , Efficiency calibration , HPGe detector , gamma-Ray , Efficiency curve , Absorption effect , Standard source , Representative point method
Journal title
Applied Radiation and Isotopes
Serial Year
2008
Journal title
Applied Radiation and Isotopes
Record number
548446
Link To Document