Title of article :
Experimental characterization and Monte Carlo simulation of Si(Li) detector efficiency by radioactive sources and PIXE
Author/Authors :
M. Mesradi، نويسنده , , A. Elanique، نويسنده , , A. Nourreddine، نويسنده , , A. Pape، نويسنده , , D. Raiser، نويسنده , , A. Sellam، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
6
From page :
780
To page :
785
Abstract :
This work relates to the study and characterization of the response function of an X-ray spectrometry system. The intrinsic efficiency of a Si(Li) detector has been simulated with the Monte Carlo codes MCNP and GEANT4 in the photon energy range of 2.6–59.5 keV. After finding it necessary to take a radiograph of the detector inside its cryostat to learn the correct dimensions, agreement within 10% between the simulations and experimental measurements with several point-like sources and PIXE results was obtained.
Keywords :
Si(Li) detector efficiency , MCNP , GEANT , PIXE , simulation , radioactive sources
Journal title :
Applied Radiation and Isotopes
Serial Year :
2008
Journal title :
Applied Radiation and Isotopes
Record number :
548447
Link To Document :
بازگشت