Title of article
Primary standardization of 67Ga radiopharmaceuticals
Author/Authors
Ming-Chen Yuan، نويسنده , , Ing-Jane Chen، نويسنده , , Chu-Fang Wang، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
5
From page
976
To page
980
Abstract
The two-dimensional extrapolation technique and dead time extrapolation technique of 4πβ-γ-coincidence counting were used to standardize the activity of 67Ga. The counting results of the two absolute counting techniques showed good agreement while the two-dimensional extrapolation technique could save about 50% of the total counting time in this study. The accuracy of the Capintec CRC-15R radionuclide calibrator used by the radiopharmacy of INER was studied in this research. A new calibration setting number, 99, was recommended to the radiopharmacy of INER in the Bayer 10 mL sterile empty vial geometry.
Keywords
67Ga , Coincidence counting , Capintec CRC-15R calibration setting. , Standardization
Journal title
Applied Radiation and Isotopes
Serial Year
2008
Journal title
Applied Radiation and Isotopes
Record number
548486