Title of article :
Monte Carlo simulations of the SNM spectra for CZT and NaI spectrometers
Author/Authors :
Liang Chen، نويسنده , , Yi-Xiang Wei، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
The detector characteristics of CZT and NaI spectrometers that affect the choice of the nuclide analysis algorithm were analyzed with a general purpose Monte Carlo code to simulate the SNM spectra. The physical and geometric conditions strictly adhered to the IAEA series test requirements for a portable radionuclide identification device. The Hecht equation was used to describe the incompletion charge collection effect for the CZT detector, and a comparison calibration to determine the CZT spectrometer equivalent noise charge. The NaI spectrometer was calibrated by experiments. The calculated results agreed well with experiments.
Keywords :
NaI detector , Nuclide analysis , CZT detector , Monte Carlo simulation , Geant4
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes