• Title of article

    An approximation for response function to γ-rays of NaI(Tl) detectors up to 1.5 MeV

  • Author/Authors

    Ahmet Cengiz، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    6
  • From page
    1371
  • To page
    1376
  • Abstract
    The response functions of a 7.62×7.62 cm NaI(Tl) scintillation detector to photons from point γ-ray sources, 10 cm from the scintillator surface, in the energy up to 1.5 MeV, were calculated using the Monte Carlo method, applying simple approximations based on the peak to total ratio and the detector resolution. The Compton continuum of the detector response function was assumed as an isotropic (rectangular) region for the photon energies up to 1 MeV. In the energies between 1 and 1.5 MeV, the Compton continuum was obtained assuming a single Compton scattering with free electrons. The photopeak of the detector response function was assumed as a line. Each determined channel of the response function was distributed to Gaussian functions. The obtained response functions were compared with the experimental values and a good agreement was found.
  • Keywords
    response function , NaI(Tl) detector , Monte Carlo method , Peak to total ratio , Detector resolution