• Title of article

    Continuous and discrete on-site detection of radon-222 in ground- and surface waters by means of an extraction module

  • Author/Authors

    Axel Schmidt، نويسنده , , Michael Schlueter، نويسنده , , Martin Melles، نويسنده , , Michael Schubert، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    6
  • From page
    1939
  • To page
    1944
  • Abstract
    A robust and straightforward method for on-site determination of radon-222 activity concentrations in water is presented. The methodical approach is based on the principle of liquid–gas-membrane extraction. An extraction module, which consists of hollow polypropylene fibres, allows radon stripping from the water of interest into a connected closed air-loop. The resulting radon-in-air activity concentration, which can easily be reconverted into the original radon-in-water activity concentration is measured by means of a standard mobile radon-in-air monitor. The experimental set-up allows radon detection in discrete water samples as well as continuous water pump streams. The technique, covering a wide activity concentration range, is in particular advantageous for measurements of radon-in-water in the field or on research cruises.
  • Keywords
    Radon-222 , On-site technique , groundwater , surface water
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2008
  • Journal title
    Applied Radiation and Isotopes
  • Record number