• Title of article

    Impact to the head increases cochlear implant reimplantation rate in children

  • Author/Authors

    Jan Bernd Weise، نويسنده , , Joachim Muller-Deile، نويسنده , , Goetz Brademann، نويسنده , , Jens Eduard Meyer، نويسنده , , Petra Ambrosch، نويسنده , , Steffen Maune، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    339
  • To page
    343
  • Abstract
    Objective: To describe the outcome of cochlear implantation in children and to discuss the cause and management of cochlear reimplantation. Study design: Retrospective chart review. Methods: The medical records of 110 patients younger than 18 years of age, who underwent cochlear implantation at the Department of ORL, Head and Neck Surgery, of the University of Kiel, Germany, were reviewed for demographics, complications, and history of revision surgery. The patients had previously had implantation with either Nucleus (including the Contour) devices or MED-EL devices. Results: Length of use before cochlear explanation ranged from 4 days to 3.9 years. Reimplantation was caused by traumatic device failure, wrong electrode insertion and infection of implanted area. Results indicated a reimplantation rate of 5.4% in children compared to 0.8% in adults, mostly resulting from the greater risk of children receiving an impact to the head. Postoperative performance data showed no decrease in scores taken before failure. Conclusions: Though young children who are developing their motor skills are probably at greater risk of a cochlear reimplantation resulting from device failure following head trauma, surgical revision with reimplantation can be performed safely and without decrement to performance.
  • Keywords
    cochlear implant , Reimplantation , children , outcome
  • Journal title
    Auris Nasus Larynx
  • Serial Year
    2005
  • Journal title
    Auris Nasus Larynx
  • Record number

    567773