    Temporal decline in defibrillation thresholds with an active pectoral lead system

  • Author/Authors

    Eric J. Rashba، نويسنده , , Mary R. Olsovsky MD، نويسنده , , Stephen R. Shorofsky، نويسنده , , Malcolm M. Kirk، نويسنده , , Robert W. Peters، نويسنده , , Michael R. Gold، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    6
  • From page
    1150
  • To page
    1155
  • Abstract
    OBJECTIVES The objective of this study was to characterize temporal changes in defibrillation thresholds (DFTs) after implantation with an active pectoral, dual-coil transvenous lead system. BACKGROUND Ventricular DFTs rise over time when monophasic waveforms are used with non-thoracotomy lead systems. This effect is attenuated when biphasic waveforms are used with transvenous lead systems; however, significant increases in DFT still occur in a minority of patients. The long-term stability of DFTs with contemporary active pectoral lead systems is unknown. METHODS This study was a prospective assessment of temporal changes in DFT using a uniform testing algorithm, shock polarity and dual-coil active pectoral lead system. Thresholds were measured at implantation, before discharge and at long-term follow-up (70 ± 40 weeks) in 50 patients. RESULTS The DFTs were 9.2 ± 5.4 J at implantation, 8.3 ± 5.8 J before discharge and 6.9 ± 3.6 J at long-term follow-up (p < 0.01 by analysis of variance; p < 0.05 for long-term follow-up vs. at implantation or before discharge). The effect was most marked in a prespecified subgroup with high implant DFTs (≥15 J). No patient developed an inadequate safety margin (<9 J) during follow-up. CONCLUSIONS The DFTs declined significantly after implantation with an active pectoral, dual-coil transvenous lead system, and no clinically significant increases in DFT were observed. Therefore, routine defibrillation testing may not be required during the first two years after implantation with this lead system, in the absence of a change in the cardiac substrate or treatment with antiarrhythmic drugs.
  • Keywords
    ANOVA , Analysis of variance , defibrillation threshold , DFT , ICD , implantable cardioverter-defibrillator
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Serial Year
    2001
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Record number

    596843