Title of article :
Application of microwave energy in cardiac tissue ablation: from in vitro analyses to clinical use
Author/Authors :
Mathew R. Williams، نويسنده , , Michael Knaut، نويسنده , , Dany Bérubé، نويسنده , , Mehmet C Oz، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
Background
Recently, there has been renewed interest in simplifying a surgical cure for atrial fibrillation. Microwave energy ablation provides an easier and faster surgical technique, in lieu of traditional “cut and sew,” which can also be applied on the beating heart.
Methods
Specific absorption rate (SAR) and thermal profiles were determined after microwave energy ablation intended for clinical application. Lesion geometry measures obtained in vitro with tissue equivalent media were compared with values collected from animal myocardium.
Results
Thermal profile analysis demonstrated lesion penetration depths of 2.5, 3, and 5 mm after 8, 16, and 24 seconds, respectively, with microwave power application of 40 W. Dosimetric studies performed using animal myocardium corresponded to the thermal profile analysis and showed that lesion depth was controlled by the output power and the ablation time.
Conclusions
Lesion geometry in tissue using microwave energy is similar to that predicted from in vitro analysis. The ablation depths and thermal profile of microwave ablation is favorable for performing atrial ablation, and this is corroborated by favorable early clinical results.
Journal title :
The Annals of Thoracic Surgery
Journal title :
The Annals of Thoracic Surgery