Title of article :
Graft occlusion after deployment of the symmetry bypass system
Author/Authors :
Oliver T. Reuthebuch، نويسنده , , Alexander Kadner، نويسنده , , Mario L. Lachat، نويسنده , , Marko I. Turina، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
Recently the Symmetry Bypass System (SJM, St. Paul, MN) became available. Now the system is frequently applied for vein-graft to aorta anastomoses in off-pump coronary artery bypass operations. This report describes a complication associated with the use of the Symmetry Bypass System (SJM) in a patient undergoing a standard off-pump coronary artery bypass procedure. A novel imaging system (SPY, Novadaq, Toronto, Canada) was applied for intraoperative assessment of graft function, and this system immediately diagnosed the occlusion of the proximal mechanical anastomosis caused by a mobile atheromatous aortic plaque.
Journal title :
The Annals of Thoracic Surgery
Journal title :
The Annals of Thoracic Surgery