Title of article :
The congenitally bicuspid aortic valve: how does it function? Why does it fail?
Author/Authors :
Francis Robicsek، نويسنده , , Mano J. Thubrikar، نويسنده , , Joseph W. Cook، نويسنده , , Brett Fowler، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
9
From page :
177
To page :
185
Abstract :
Background The study was intended to investigate the reason why congenitally bicuspid valves (CBAVs), which may function hemodynamically and clinically well, have a high early failure rate. Methods Observations were made on cryopreserved, then thawed human aortic roots containing CBAVs. Valvular function was studied in the left heart simulator using conventional and 500-frames/second cinematography, intravascular ultrasound, by preparation of silicone molds, and by computerized digital modeling. Results The function of the clinically “normal” CBAVs is characterized by (a) excessive folding and creasing, which (unlike in the trileaflet valve) persist throughout the cardiac cycle; (b) extended areas of leaflet contact; (c) significant morphologic stenosis; and (d) asymmetrical flow patterns and turbulence. Conclusions The above features subject the CBAV to abnormally high stresses and may lead to early thickening and eventually calcification and stenosis. The abnormal flow patterns also predilect for dilatation and dissection of the ascending aorta.
Journal title :
The Annals of Thoracic Surgery
Serial Year :
2004
Journal title :
The Annals of Thoracic Surgery
Record number :
607218
Link To Document :
بازگشت