Title of article :
Beating-heart epicardial radiofrequency ablation: optimal temperature setting
Author/Authors :
Yoshito Inoue، نويسنده , , Issei Kiso، نويسنده , , Ryuichi Takahashi، نويسنده , , Atsuo Mori، نويسنده , , Kenta Motogami، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
Purpose
Pulmonary vein isolation is a simple procedure, which has recently been reported as an effective treatment for the termination of atrial fibrillation. Although there are several clinical reports of beating-heart epicardial ablation, the optimal temperature has not been experimentally investigated. We evaluated the effective temperature for the placement of circular lesions around the pulmonary vein–left atrial junction.
Description
Twelve swine underwent epicardial ablation to create linear conduction block lesions around the pulmonary vein–left atrial junction by a seven-electrode ablation catheter. The ablation was performed at 60°C (group I), 70°C (group II), 80°C (group III), and 90°C (group IV) for 120 seconds. The creation of a firm conduction block across the ablated lesion under pacing was compared.
Evaluation
Complete conduction block was observed in all groups except group I. However, heat injury to adjacent structures in group IV and transient discoloration of the tissue surrounding coronary arteries in groups III and IV were observed.
Conclusions
The effective temperature for epicardial radiofrequency pulmonary vein isolation was 120 minutes and above 70°C.
Journal title :
The Annals of Thoracic Surgery
Journal title :
The Annals of Thoracic Surgery