Title of article
Severe Stenosis of Anastomoses by Using the Symmetry Aortic Connector System
Author/Authors
Jose M. Melero، نويسنده , , Carlos Porras، نويسنده , , Miguel Such، نويسنده , , Eduardo Olalla، نويسنده , , Juan Alonso-Briales، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
3
From page
1831
To page
1833
Abstract
The St. Jude Medical Symmetry Aortic Connector System was developed to create the proximal vein graft anastomoses in coronary artery bypass grafting. We describe three symptomatic patients with severe stenosis of the proximal anastomosis several months after using the Symmetry aortic connector system. Intravascular ultrasound study showed anastomotic neointimal hyperplasia.
Journal title
The Annals of Thoracic Surgery
Serial Year
2004
Journal title
The Annals of Thoracic Surgery
Record number
608101
Link To Document